Product Overview
BluVac+ LTE User’s Guide
5
Battery Installation & Replacement
1. If installed, remove the protective boot.
2. Remove the battery cover from the rear of the BluVac+ LTE by compressing tab at the base of
the battery cover.
3. If necessary, remove and detach old battery from battery clip.
4. Attach battery clip to new battery and insert into battery compartment. Replace battery cover by
aligning tab and snapping back into place.
5. Reinstall the protective boot.
IMPORTANT: TO PREVENT DAMAGE FROM LEAKING BATTERIES, DO NOT LEAVE A DEAD
BATTERY INSIDE THE BluVac+ LTE. REMOVE BATTERY IF THE BluVac+ LTE IS NOT TO BE
USED FOR AN EXTENDED PERIOD OF TIME.
Battery Level Indicator
The battery level indicator shows the relative strength of the battery. Four bars indicate full power.
As the battery is depleted, the number of bars displayed decreases, until no bars are left. At this
point, it is necessary to replace the batteries (refer to the Battery Installation section above). If the
battery power drops to the point where the BluVac+ LTE can no longer function accurately, the alarm
will beep 10 times and the power will turn off automatically. When the battery is replaced, and the
BluVac+ LTE is turned back on, the BluVac+ LTE will resume operation with all previous settings
intact.
Hi-Pressure Indication
If the sensed vacuum pressure exceeds 25,000 Microns (3333.1 Pa, 33.331 mBar, 25,000 mTorr, or
25.000 mmHg), the display will show .
Units
Change the displayed units by pressing the key on the keypad. The appropriate units indicator
on the LCD display will show Pa, mBar, mTorr, mmHg, or Microns depending on the units set.
Sleep Mode
Unlike other Digital Vacuum Gauges, the BluVac+ LTE helps to conserve battery life through
advanced power management. After 5 minutes of displaying , the BluVac+ LTE will enter “Sleep
Mode” in which the sensor and backlight are turned off, thus reducing the load on the battery and
extending the battery life up to 300 hours or more. While in this mode, the display will show .
Approximately every 35 seconds, the gauge will automatically check the pressure and exit Sleep
Mode if the pressure is less than 25,000 Microns. The gauge can be manually brought out of Sleep
Mode by pressing any key on the keypad. While in Sleep, the gauge will beep twice every 5 minutes
to remind you that it is on. The BluVac+ LTE will not sleep in the Calibration Mode (see the
Calibration section below), or if the wireless function is activated and a connection is active.
Auto Power-Off
After 1 hour in “Sleep Mode” the BluVac+ LTE will automatically turn itself off to further conserve
battery power.