Product Overview

BluVac+ Professional User’s Guide
10
8. While the pressure display is flashing, press and/or hold the or keys to increase or
decrease the RTT Target Pressure. When compete, press and release the key.
9. While the minutes display is flashing, press and/or hold the or keys to increase or decrease
the RTT Target Time minutes. When compete, press and release the key.
10. While the seconds display is flashing, press and/or hold the or keys to increase or
decrease the RTT Target Time seconds. When compete, press and release the key.
11. The gauge will beep three times (if sound is enabled), and return to normal mode. The new
program is now saved.
12. At any point in the Program Set Mode, pressing and holding the key will result in saving the
current program and returning to normal mode.
13. Once programmed, the BluVac+ Prowill remember the settings until changed again, even if the
power is turned off or the battery removed.
NOTE: The units cannot be changed while in the Set Mode.
Executing a Program (Run Mode)
Press and release . The indicator will illuminate and (EVAC) or (RISE), indicating
the current program execution mode, will flash on the main display.
Press the key to select the desired execution mode. The main display will toggle between
and with each press of the key.
Executing the Evacuation Program
1. Ensure is flashing on the display. Press to select , as above, if not.
2. Press to start the program.
3. The RUN and TIME indicators will illuminate on the display. RUN will flash to indicate the program
is running.
4. Start the vacuum pump.
5. When the vacuum pressure drops to less than the target evacuation pressure, the timer will
start. Subsequently, if the pressure exceeds the target evacuation pressure, the timer will stop
and resume counting when the pressure again drops to less than the target evacuation
pressure.
6. When the timer expires, the BluVac+ Pro will sound an alarm, flash the backlight, and blink the
DONE indicator. Press any key to silence the alarm. The program is complete.
7. Blank-off and stop the vacuum pump.
Executing the Rise Time Test Program
1. Ensure the vacuum pump is blanked-off and stopped.
2. Ensure is flashing on the display. Use the to select , as above, if not.
3. Press to start the program.
4. The RUN and RISE TIME indicators will illuminate on the display. RUN will flash to indicate the
program is running.
5. Immediately, the BluVac+ Pro will begin monitoring the rise rate of pressure, though the clock
will remain stopped until the pressure exceeds the Evacuation Target Pressure.
6. If a determination is made that the pressure will not exceed the Rise Time Pressure Target
within the allotted time, the BluVac will alarm and indicate regardless of the time remaining.
The analysis is performed, and a can be achieved, even if the clock is not running (i.e. the
Evacuation Target Pressure has not yet been achieved.