User Guide

DS-220 User’s Guide
14
Battery Level Indicator
The battery level indicator shows the relative strength of the batteries. Full power is indicated by
four bars. As the batteries are depleted, the number of bars displayed decreases, until no bars
are left. At this point, it is necessary to replace the batteries (refer to the Battery Installation
section above). If the battery power drops to the point where the DS-220 can no longer function
accurately, the alarm will beep 20 times and the power will turn off automatically. When the
batteries are replaced, and the scale is turned back on, the DS-220 will resume operation with all
previous setting intact.
Calibration
The DS-220 is calibrated at the factory to provide high accuracy. While recalibration should
seldom, if ever, be necessary, recalibration is performed easily without sending the product to the
manufacturer for service. A known calibration weight (in kilograms) is required to recalibrate the
scale, and can be any value between 5 kg and 100 kg. 25 kg is recommend. Best accuracy is
obtained by using the largest weight available (i.e. a 50 kg calibration weight will provide 10x
better calibration accuracy than a 5 kg calibration weight). In addition, it is necessary to know the
weight to within 2 grams.
To recalibrate the DS-220, perform the following steps:
1. Place the DS-220 weighing platform on a firm, level surface.
2. Turn the power to the DS-220 off.
3. Remove all objects from the platform.
4. While holding , press and release . The DS-220 will turn on, and after a few seconds
the display will show SET CAL and 25.000 kg.
5. Using or adjust the displayed value (in 1g increments) so that it is equal to the
weight of the reference calibration weight. Use the key to adjust in 1 kg increments.
6. Press . A number representing the unloaded platform will appear on the display. Press
to lock in this value. The display will read zero.
7. Apply the calibration weight to the platform. A number representing the applied reference will
appear on the display. Allow at least 10 seconds for this number to stabilize.
8. Press . The DS-220 is now recalibrated to the reference weight.
9. Refer to the Quick Start section above for setting the desired Mode, Units, and Resolution.