Instruction manual

9
Readheads
ABS style readheads have “BLACK END OF SCALE” labels on the
cover, and the telephone-style wire exits from the corner of the housing.
Extreme care must be taken not to damage the six brass “fingers” inside
the readhead housing. ABS style readheads, used in all Model 150 and
Model 250 systems, must be mounted on ABS scales with a particular
orientation. Each readhead has an arrow on the label pointing in the
direction of the “BLACK END OF SCALE” (each ABS style scale will
have one end painted black). This relationship is very important, since
the readhead will work, but produce erratic results if incorrectly installed.
To insure proper operation, be sure the arrow on the readhead is
pointing toward the BLACK end of the scale. The standard readhead
has 3m (10 ft.) of cable. For special cable lengths, contact Accurate
Technology, Inc. The ABS style readheads are standard in the Model
150 and Model 250 systems.
Incremental style readheads do not have labels on the cover, and the
telephone style wire exits from the center of the housing. Extreme care
must be taken not to damage the twelve brass “fingers” inside the
readhead housing. Incremental readheads may be mounted on
incremental scales with any orientation. Incremental readheads are
used in ProStop, ProKit, Measuring Kits, ProCaliper, and ProPanel.