Installation manual

Manual Part # 800-1402-001 Rev E Page 39 of 48
The Digital Readout Programming Parameters are listed below.
Values in [ ] are the range of values available for that Parameter.
Factory defaults values are shown in Bold Red.
NOTE: Programming parameters are not sequentially numbered due to
firmware differences between systems and provision for future
enhancements / features.
Pr 1 – Datum Key [0 to + 999.999in] or [0 to +9999.99mm]
The programmed value that will be recalled whenever the Datum key is
pressed during normal operation.
Default = 0.00
Pr 2 – Direction of Travel [0 or 1]
This parameter controls the sign of travel (positive vs. negative) when the
measuring system is moved.
Default = 0
Pr 3 – Key Lockout [0 or 1]
This parameter controls the operation of the +, - and DATUM keys. If enabled,
(set to 1), these keys will not function and the LOCK
LOCKLOCK
LOCK symbol will appear on the
display. This prevents accidental changes when depressing these keys during
normal operation.
Default = 0
Pr 4 – Display Resolution [1, 2, 3 or 4]
This parameter sets the number of places to the right of the decimal point on
the display. When the DRO is in a decimal mode (in, mm or cm), it will auto-
range to the next resolution if the value is too large to be displayed in the
current resolution but is displayable in an alternate resolution.
A value of 1 will display x.x.
A value of 2 will display x.xx
A value of 3 will display x.xxx
A value of 4 will display x.xxxx
Default = 3
NOTES:
Decimal inches have a maximum of 3 decimal places (4 places + .0005).
Millimeters have a maximum of 2 decimal places (even if the programming
parameter is set to 3 or 4).
This programming option has no effect when displaying fractions.
Pr 5 – Movement Required for DRO Auto-On function [0.3 to 10mm]
This parameter sets the amount of encoder, or system, movement required to
automatically wake up the digital readout when it is turned off.
Default = 0.1mm