User's Manual Part 6
ProScale M150, M250 and all products with General Purpose LCD Digital Displays 34 of 34
Pr8 – Automatic Power Off
[0 to 60]
Default = 15
.
Sets the amount of time in ‘minutes without activity’ before the display
automatically turns off.
0 = Disables Auto Off.
Encoder motion or ON/OFF key “wake-ups” the display and resets the timer.
Pr9 – Auxiliary Keys Enable/Disable
[0..
7
]
0 = ABS/INC, MON and HOLD Disabled
1 = ABS/INC Key Enabled
2 = MON Key Enabled
4 = HOLD Key Enabled
7 = All Keys Enabled
To enable keys, add up combination of key values. A value of 2 enables only the
MON key. A value of 7 enables all 3 Keys.
Pr10 – Offset Addition Enable
[
0
, 1]
0 = Offset Addition Disabled
1 = Offset Addition Enabled.
SEE ALSO Pr11, Pr12, Pr13
Pr11 – Offset Addition 1
[-999.999 to 999.999in] or [-9999.99 to 9999.99mm]
When offset 1 is selected (see section 4 -
Offset Addition
), this value is added to
the current ABS position.
Default = 1.000IN
Only active if Pr10 is set to 1.
Note: Default is set in Inches
Pr12 – Offset Addition 2
[-999.999 to 999.999in] or [-9999.99 to 9999.99mm]
When offset 2 is selected (see section 4 -
Offset Addition
), this value is added to
the current ABS position.
Default: 1.500IN
Only active if Pr10 is set to 1.
Note
:
Default is set in Inches
Pr13 – Offset Addition 3
[-999.999 to 999.999in] or [-9999.99 to 9999.99mm]
When offset 3 is selected (see section 4 -
Offset Addition
), this value is added to
the current ABS position and displayed.
Default: 2.000IN
Only active if Pr10 is set to 1.
Note
:
Default is set in Inches
Pr14 – Output Signal Mode
[
0
, 1]
Configures the hardware output signal for activation on MONitor drift conditions or
Upper/Lower limit alarm conditions. (24VDC displays only)
0 = Monitor drift,
1 = Limit.Alarm
SEE ALSO Pr15, Pr16, Pr17