Specifications

8
88
8
DigiPump
DigiPump DigiPump
DigiPump SR7x
SR7xSR7x
SR7x Sy
Sy Sy
Syringe Pump
ringe Pump ringe Pump
ringe Pump Operator
OperatorOperator
Operators
ss
s Manual
Manual Manual
Manual
(3) Before the installation, manually remove all the air inside the
syringe and then place the
syringe on the mounting grooves.
(4) Return the Syringe Holder to original position compressing and holding the syringe in place.
When the Syringe Holder compresses the syringe, the syringe size indicator is displayed in the
screen.
Step 4 Set the Infusion Rate
With the DigiPump SR7x in STOP mode, rotate the adjusting Knob to set the desired flow
rate and press the adjusting Knob to confirm and save the adjusted flow rate.
Note: Turn the adjusting Knob clockwise to increase the flow rate or anti-clockwise to
decrease the flow rate.
Step 5 Set Bed Number (Only when using together with Infusion Supervision System)
Press SELECT key. The display light indicator changes to Bed No. Rotate the adjusting Knob
to set the desired Bed No and press the adjusting Knob to confirm and save.
Note: Turn the adjusting Knob clockwise to increase the Bed No or anti-clockwise to
decrease the Bed No.
Step 6 Clear Accumulated Volume
(1) Press SELECT key. The display light indicator changes to Bed No., then press the SELECT
key again and the display light indicator changes to Ʃ ml (accumulated volume).
(2) Press CLEAR key, the accumulated volume will clear to zero ml ‘0’.