pump infuses at maximum rate for the syringe.
17
Adjust & Confirm Knob
Rotating the knob adjust value. Pressing the knob
confirm and save the adjusted value.
18
Clutch button
Keep pressing the button to move the piston freely.
When unfastening the button, the clutch will gear,
and can only be moved by the engine.
19
Piston
Pushes the syringe during infusion.
20
50 ml syringe indicator light Light is ON when the 50ml syringe is placed
properly.
21
20 ml syringe indicator light Light is ON when the 20ml syringe is placed
properly.
22
10 ml syringe indicator light Light is ON when the 10ml syringe is placed
properly.
23
Syringe Holder Hold and detect the syringe.
24
User-friendly instructions Operation guide for quick reference.
1.9 Battery Charge
1. As shown in figure 1, after the pump is connected to the AC power and turned on, the AC
indicator light is on and the battery begins to charge automatically. When the battery is fully
charged, it will stop charge automatically.
2. As shown in figure 2, the AC indicator light will be off if the AC power is disconnected. The
syringe pump is running on battery and the battery indicator is showing full capacity. As the battery
capacity decreases the lights turns-off from top to bottom.
Note: It requires 14 hours to fully charge the battery after it is fully depleted.
Figure 1 Figure 2
1.10 Pump Installation
1. The syringe pump is used in horizontal position.
2. Make sure the screw center with thread hole in the enclosure, then rotate the screw to fasten the
clip on the syringe pump.
3. Check the stability of the IV stand.
4. Rotate the second clip screw to fasten the syringe pump on the IV stand.