7. Warning: when using Non standard syringe or setting syringe parameters incorrect, the accuracy
may not be maintained and the discrepancy may reach 5% or above. When changing Syringe
types and syringe calibration settings, it is the user responsibility to ensure that required
accuracy is obtained.
8. Warning: the used syringe shall not be installed on other infusion devices to avoid potential risk.
9. Warning: Keep cell phone or other high frequency power devices away from the syringe pump
within 2’ (0.6m).
10. Warning: There is no air in line detection sensor in the pump. Make sure that there is no air
bubble in the syringe and the infusion set before using.
11. Warning: Stop using the machine when there is alarm.
12. Warning: Do not use the syringe pump in places where flammability and explosion can occur.
1.4 Product functions and features
(1) Precise control of infusion rate.
(2) Precise control of infusion volume.
(3) Smooth flow.
(4) Audible and visual alarms for infusion completion, near completion, occlusion, low battery,
syringe abnormal, control abnormal and no AC power.
(5) Advanced infusion control.
1.5 Technical Specifications
Product name DigiPump SR7x
Maximum flow rate 500ml/h (Different syringes have different
maximum rate.)
Flow rate range 10ml syringe: 0.1 - 100ml/h
20ml syringe: 0.1 - 200ml/h
50ml syringe: 0.1 - 500ml/h
Infusion increment 0.1ml/h
Bolus rate 10ml syringe: 100ml/h
20ml syringe: 200ml/h
50ml syringe: 500ml/h
Accumulated infusion volume 0.1 – 999.9ml
Accuracy +/- 3% (up to +/- 1% with syringe calibration)
Power supply AC 100~240V,50/60Hz, 25VA.
Battery Rechargeable lithium battery, 7.4 V, 1650mAh.
Maximum power consumption 25VA, running more than 5 hours after being
fully charged.
Battery charge When syringe pump is connected to AC power,
the SR7x need to be ON and the battery will
automatically recharge. (about 8~14 hours to
recharge fully.)
Displayed information Flow rate, accumulated infusion volume, syringe
size, battery capacity, bed No., AC power
indicator.
Alarm information NEAR completion, OVER, OCCLUSION, low
battery, abnormal 1 (communications failure),