Syringe
Disengaged
1.The syringe is
disengaged or
not properly
installed.
2.The syringe’s
parameter is
wrongly set.
3. Something is
wrong with
the syringe
sensor
No.1:Install the syringe
again
No.2:Refer to item
3.2
Syringe Configuration
Change
No.3:Contact Digicare
Technical Support
Control
Abnormal
1. The syringe’s
parameter is
wrongly set
2. the syringe
pump’s
motor can’t
run properly
No.1:Refer to 3.2 Syringe
Configuration Change
No.2 : Contact Digicare
Technical Support
5.2 Troubleshooting
Description Result Cause Solutions
Press START key
but the light for
LED operation is
not on.
The syringe
pump does not
work.
The syringe is not properly
installed.
The syringe configuration
parameter is not rightly set.
Reinstall the syringe. Check
if the right indicator light is
on. Or configure the syringe
settings.
Turn the knob but
infusion rate does
not change.
The infusion rate
does not change.
The rate cannot be changed
during infusion operation.
Press STOP key to stop the
syringe pump. Turn the knob
to set the needed value and
press the knob to save the
value.
The syringe pump
can’t turn-on with
the battery
The syringe
pump can’t turn
on
Bad battery. The syringe
pump can not be used before
it is connected to A.C.
electric supply.
Connect the syringe pump to
A.C. power supply before
operation (Make sure the
pump is ON).
The occlusion
alarm signal
sounds soon after
the syringe pump
starts operation.
The occlusion
light is on and
the syringe
pump stops.
The occlusion sensitivity
value of the syringe pump is
not properly set.
Reset the occlusion
sensitivity value.
The syringe size
sensor can’t detect
the syringe
The indicator
light of the
corresponding
syringe is not on
The compressed strut is not
tightly pressed on the
syringe. Or the syringe
setting is not rightly set.
Pull the compressed strut of
the syringe up, reinstall the
syringe and press it onto the
syringe tightly. Or reset the
syringe setting as 3.2