Specifications
15
1515
15
DigiPump
DigiPump DigiPump
DigiPump SR7x
SR7xSR7x
SR7x Sy
Sy Sy
Syringe Pump
ringe Pump ringe Pump
ringe Pump Operator
OperatorOperator
Operator’s
ss
s Manual
Manual Manual
Manual
4.3 Preventive Inspection
1. Flow Rate accuracy check.
Use one volume calibrated recipient to verify Flow Rate and Volume accuracy every time
new syringes are being used or every six months.
2.Battery Inspection
Batteries have limited life time. Replace the batteries when they are not holding charge.
If replacing batteries are required, please contact Digicare. Model: Lithium Polymer
battery (7.4V,1650mAh).
Use the pump running in its internal battery at least once a month till its depletion and
the power of the pump turns off automatically, for insuring battery performance and
prolonging its life. After battery depletion, recharge 8~14 hours for next use.
The following inspection need to be taken every six months.
(1) Connect to AC power, about 8 hours for recharging.
(2) Turn on the syringe pump and install 50ml syringe.
(3) Set up infusion rate: 25ml/h and start syringe.
(4) Observe operation of the syringe pump till it turns off based on low battery.
If the syringe pump takes 90 minutes or even longer time from starting
infusion to shutdown, the battery is in good condition.
If the syringe pump takes 45-90 minutes from starting infusion to shutdown,
the battery life close to its end.
If the syringe pump takes less than 45 minutes from start infusion to shutdown,
the battery is up to its end and recharge of battery is required.
(5) After battery inspection, recharge the battery again for usage next time.
3.Routine Maintenance
intervening time Routine maintenance procedures
According to hospital policy
Thoroughly cleaning of syringe pump case is
required before or after long period of storage.
At least once a year for inspection
1.Inspection of AC power cord and wires.
4.4 Environmental Protection
The life time of the products is 3 years. Machines over its life should be discarded, please
contact the manufacturer or distributor for more relevant information.
1. DigiPump SR7x syringe pumps that are no longer in use could be delivered to its
distributors or manufacturer for proper recycling.
2. Used Lithium Polymer batteries could be delivered to its distributors or manufacturers for
handling, or handling it as per applicable laws and regulations.