3.3 Occlusion Sensibility Adjust
The occlusion pressure is sensibility can be adjusted by the user. The value range goes from 12 to 20.
The sensibility is set in the factory to 13. The smaller the value is, the more sensitive the occlusion
alarm is. The actual pressure value depends on the syringes size and conditions in the range from 10
to 25PSI.
To change the occlusion pressure sensibility value, with the syringe pump OFF, press and hold the
START + POWER keys simultaneously. The syringe pump is turned-on and a configuration mode
sound indicates the syringe pump enters configuration mode. Release all the keys together.
Note: During infusion, when the OCCLUSION alarm is triggered the syringe pump enters KVO
mode. It returns to original infusion rate when the occlusion is removed
4. Cleaning, Storage, Transport and Maintenance
4.1 Cleaning
1Keep the pump clean. Use a piece of cloth to wipe the syringe pump clean when any liquid
drops on the pump.
2. Special attention must be paid when cleaning the syringe pump. Use a tampon moistened
with 75% of alcohol to wipe the outer case.
3Make sure the AC power is disconnected and the pump is turned off.
4Do not clean the syringe pump by using xylene, acetone or similar corrosive cleaning
products to avoid outer case damage.
The above mentioned operation is just for guidance. More proper methods should be used for
sanitizing effect inspection.
4.2 Transport and Storage
Transport and Storage environmental conditions: temperature: -15C50Catmosphere
pressure: 86kPa106kParelative humidity: 20% ~ 90%.