2)- 20ml syringe’s parameter setting: With the syringe pump OFF, press and hold the BOLUS +
POWER keys simultaneously. The syringe pump is turned-on and a configuration mode sound
indicates the syringe pump enters configuration mode. Release all the keys together. The 20ml
indicator light will be ON and the 20ml current parameter value is displayed. Turn the adjusting
knob to change the value to the one measured in step 1, then press the knob to save the new value.
3)- 10ml syringe’s parameter setting: With the syringe pump OFF, press and hold the STOP +
POWER keys simultaneously. The syringe pump is turned-on and a configuration mode sound
indicates the syringe pump enters configuration mode. Release all the keys together. The 10ml
indicator light will be ON and the 10ml current parameter value is displayed. Turn the adjusting
knob to change the value to the one measured in step 1, then press the knob to save the new value.