Step 1 Measure the syringe length
Measure the new syringe length in millimeters (mm).
Step 2 Install the syringe
Turn off the syringe pump. Install the new syringe on the pump. Push the syringe piston to
compressing the syringe all the way to the syringe’s end.
Warning: The correct installation of the syringe and the piston is essential when changing the
calibration settings or error may occur.
Step 3 Set the syringe parameter
1)- 50ml syringe’s parameter setting: With the syringe pump OFF, press and hold the CLEAR +
POWER keys simultaneously. The syringe pump is turned-on and a configuration mode sound
indicates the syringe pump enters configuration mode. Release all the keys together. The 50ml
indicator light will be ON and the 50ml current parameter value is displayed. Turn the adjusting
knob to change the value to the one measured in step 1, then press the knob to save the new value.