3. Configuration
3.1 Enable and Disable ‘NEAR’ Alarm Function
(1) With the syringe pump off, press and hold the START + CLEAR + POWER keys
simultaneously. The syringe pump is turned-on and a configuration mode sound indicates
the syringe pump enters configuration mode. Release all the keys together.
(2) The syringe pump enters into the configuration mode to lock and unlock the NEAR alarm.
The NEAR alarm light is turned on and the screen display a value. 01 stands for NEAR
alarm function DISABLED and 02 stands for NEAR alarm function ENABLED.
3.2 Syringe Configuration Change
The DigiPump SR7x syringe pump is calibrated in the factory to the following disposable
syringes:
Sizes for Reference in (mm)
Manufacturer
10ml 20ml 50ml
BD 61.0 70.0 89.5
Length from Size indicator marking to 0ml
e.g.:50ml syringe