As shown in the figure bellow, when the infusion is completed, the OVER indicator light is
ON, accompanied with an audible alarm signal to alert the user. Press the STOP key and the
syringe pump stops operation.
Note: The OVER alarm can be silenced by pressing the adjusting knob. The OVER audible
alarm will sound again in 2 minutes.
Step 9 Turn off the power
Press the POWER key for 3-5 seconds and then release it, the machine power is turned-off.
2.2 BOLUS Function
During the infusion process, if bolus infusion is needed, while the on-going flow rate is lower
than the maximum rate for the syringe size in use, keep and hold the BOLUS key, the machine
will infuse at maximum rate for the syringe in use. The Infusion Rate returns to original rate
after releasing the BOLUS key.
Syringe Size Bolus rate
10 ml 100 ml/h
20 ml 200 ml/h
50 ml 500 ml/h