Instruction Manual
R3-081016       3 
Configure parameters   
  Refer to the data management software instructions for details (pages 8 –14) 
Default Settings   
Other settings   
Logging interval   
5 minutes 
0-11 hours; 0-59 minutes; 0-50 seconds   
Stop mode:   
Software 
Manual stop (see pp. 3-4 )   
Start Mode:   
Instant on  
Manual and timing start (pg. 3)   
Temp. Unit  
o
C 
o
F 
Buffer Ring:   
Disabled 
Enabled  
Alarm 
Settings 
H1 (high alarm) =  
10 
o
C 
User defined   
L1 (low alarm) =  
2 
o
C 
Alarm Mode:  
Single 
Cumulative   
Start the data logger 
There are three modes to start the logger – instant-on, manual start, and timing start. 
1.  Instant-on:  After  parameter  configuration,  the  data  logger  starts  recording 
immediately when it is disconnected from the USB. 
2.  Manual  start:  After  parameter  configuration,  press  and  hold  the  button  for  5 
seconds to start the data logger. In this mode, it has a start delay function. If this 
function is enabled, the data logger will not record data immediately after start-up 
but start recording after the set delay time elapses. 
3.  Timing start: After parameter configuration and disconnection from USB, the data 
logger starts recording when it reaches the set time. 
View data instantly   
If you need to view simple statistical information, you may directly press the button to turn 
the page and check. The LCD screen can display MKT, average value, Max value and Min 
value. Mean Kinetic Temperature (MKT) is a simplified way of expressing the overall effect 
of temperature fluctuations during storage or shipment of perishable goods. 
If you need detailed information, please connect the data logger to the computer’s USB. 
After about 3 minutes, the data will be saved. You can open it as an Excel, AI or PDF 
report.   
Moreover, you can connect the data logger to a computer and analyze the data using the 
data management software. 
Stop the data logger 
There  are  several  modes  to  stop  it  –  manual  stop,  stop  via  software,  over-Max- 
record-capacity stop (enable/disable manual stop). 
1.  Manual stop: When the data logger is recording in this mode, you may press and hold 
the button for 5 seconds to stop it. 
2.  Stop via software: You can stop the data logger via software 
by 
selecting the Stop recording option on the software platform.   
If  the  record  capacity  reaches  the  Max  value  (32,000  points) 
Caution! 
Make sure you have saved selected parameters. 