Use and Care Guide
ManualsBrandsAccuCold ManualsCompact Refrigerators19 Inch Compact Top Freezer Refrigerator with Combination Lock, Buffer Temperature Probe, Adjustable Shelves, Interior Light, NIST Calibrated Thermometers, Door Storage, Cycle Defrost, CFC Free, 2.9 cu. ft. Capacity and ADA Compliant: White, ADA Compliant
13 
Thermometer/alarm instructions (cont’d): 
Function of Controls: 
MODE/RESET Key: Press ONCE to reset Max/Min temperature record. When the unit 
is sounding an alarm, press ONCE to turn off the beeping sound. 
HIGH/LOW Key: Press ONCE to display high alarm setting. Press TWICE to display 
low alarm setting. Press THREE TIMES to display current temperature reading. 
ºF/ºC Key: Press this button to alternate the temperature display scale between 
Fahrenheit (ºF) and Celsius (ºC). 
Battery Installation: Remove battery compartment cover located on the rear of the 
unit and remove insulation strip from battery compartment. Following the polarity 
(+/-) diagram in the battery compartment, install the battery. Replace the battery 
cover. The unit will beep and all segments of the LCD will be activated. 
Programming the Alarms: 
The thermometer displays the internal temperature of the unit being monitored, as 
well as the maximum and minimum temperatures reached. The maximum and 
minimum readings reflect the highest and lowest temperatures since the battery was 
inserted or since the MODE/RESET key was last pressed. If the temperature rises 
above or falls below the set temperature range, the alarm will sound. By default, the 
high alarm temperature setting is 8°C (46.4°F) and the low alarm temperature setting 
is 2°C (35.6°F). To reset these limits, follow these instructions: 
High Alarm Temperature Setting: Press the HIGH/LOW key ONCE, then press 
MODE/RESET to enter the High Alarm temperature setting mode. Use the Up and 
Down arrows to adjust the temperature accordingly. Press MODE/RESET once 
again to confirm the setting. 
Low Alarm Temperature Setting: Press the HIGH/LOW key TWICE, then press 
MODE/RESET to enter the Low Alarm temperature setting mode. Use the Up and 
Down arrows to adjust the temperature accordingly. Press MODE/RESET ONCE to 
confirm the setting. 
By default, the low and high alarms should be audible. To turn alarms on/off: 
Audible High Alarm Setting: To make sure the high alarm is audible, press HIGH/LOW 
key ONCE (to activate HI-SET on the screen). Then press the ALERT ON/OFF key 
ONCE to activate the HIGH-ALARM beeping sound. Press again to deactivate. 
Audible Low Alarm Setting: To make sure that the low temperature alarm is audible, 
press the HIGH/LOW key TWICE to activate LOW-SET on the screen. Then press 
the ALERT ON/OFF key ONCE to activate the LO-ALARM beeping sound. Press 
again to deactivate. 
NOTE: Set high and low alarm temperature limits above and below the refrigerator/ 
freezer operating temperature to prevent false alarms.  
NOTE: For an activated high alarm and low alarm buzzer, the display should 
indicate HI-ALARM and LO-ALARM (see illustration on previous page). 
More details about the features and usage of the Traceable Thermometer/Alarm 
can be found in the User Manual accompanying the unit. 