Data Logger User Manual
AccuCold, Div of Felix Storch, Inc. | R005-07102018                                                                           
5 
Programming the Data Logger (continued) 
Low Alarm 
Temperature Setting 
From the main display screen:   
Hold the SET key for 3 seconds. Enter the correct password then press the SET key 2x to 
enter  the  LO  Temp  Alarm  setting  mode.  Use  the  up  and  down  arrows  to  adjust  the 
temperature accordingly.    Press SET key ONCE to confirm the settings. 
Alarm Hysteresis  
Hysteresis is the tolerance band to prevent alarm chattering. For instance, if the high alarm 
temperature is set at 8 
o
C with a hysteresis of 1 
o
C, once the alarm is activated, it will not 
return to normal until the temperature goes below 7 
o
C. By default, the alarm hysteresis is 
set at 1 
o
C. To reset, follow instructions below.   
From the main display screen:   
Hold the SET key for 3 seconds. Enter the correct password then press the SET key 3x to 
enter  the  Alarm  Hysteresis  setting  mode.  Use  the  up  and  down  arrows  to  adjust  the 
temperature accordingly.    Press SET key ONCE to confirm the settings. 
Alarm Delay     
The alarm delay is used to avoid unnecessary alarms when the temperature exceeds the set 
high and low alarm limits. This feature will delay alarm activation by the amount of time 
entered. By default, the alarm delay is set at 0 minutes.   To reset, follow instructions below.    
From the main display screen:   
Hold the SET key for 3 seconds. Enter the correct password then press the SET key 4x to 
enter  the  Alarm  Delay  setting  mode.  Use  the  up  and  down  arrows  to  adjust  the  time 
accordingly.    Press SET key ONCE to confirm the settings. 
Start Delay     
From the main display screen:   
Hold the SET key for 3 seconds. Enter the correct password then press the SET key 5x to 
enter  the  Start  Delay  setting  mode.  Use  the  up  and  down  arrows  to  adjust  the  time 
accordingly.    Press SET key ONCE to confirm the settings. 
Temperature Unit     
From the main display screen:   
Hold the SET key for 3 seconds. Enter the correct password then press the SET key 6x to 
enter  the  Temperature unit  setting  mode.  Use  the  up  and  down  arrows  to  adjust  the 
temperature units accordingly.    Press SET key ONCE to confirm the settings. 
When the temperature is lower than 
(High Alarm Temperature setting - 
Alarm Hysteresis), it will exit high 
temperature alarm. 
When the temperature is higher 
than (Low Alarm Temperature 
setting + Alarm Hysteresis) it will 
exit low temperature alarm.   
When an alarm condition occurs, HI-ALARM and LO-ALARM icons will appear on the display 
along with a beeping sound to alert user. The beeping sound will stay ON until the unit gets back in 
range. Press any key ONCE to stop the beeping sound.   
The high and low indicators will remain on display even when the unit gets back in range. Press   
for 3 seconds to clear the HI and LO alarm icons. 
*Note- The HI and LO alarm icons will only clear when the unit is back in range.* 