User guide
INSULIN PUMP
WARNING / ERROR
For every WARNING and ERROR it is
recommended that you consult your User
Guide for additional detailed information!
Always press twice f to snooze and
conrm the signal.
Not available in STANDARD
MENU
BRP
= Basal Rate Prole 2 – 4
f
f
BRP 4
BRP 3
BRP 2
BRP 4
BRP 3
BRP 2
RUN
STOP
Press j d to move forwards
Press j f to conrm or save
Press j a to select up
Press j d + a simultaneously
to move backwards
Press j s to select down
Press j a for the Backlight
Press j d + s simultaneously
to unlock the Keylock
Quick
Reference
Guide
05331803001/C
ACCU-CHEK and ACCU-CHEK SPIRIT COMBO are
trademarks of Roche.
Roche Diabetes Care AG
3401 Burgdorf, Switzerland
Local representatives:
AT +43 1 27787-596
AU 1800 633 457
BE 0800 93 626
CA 1-800-280-7801
CH 0800 11 00 11
CZ +420 724 133 301
DE 0180 2 000 412
DK +45 36 39 99 54
ES 900 400 000
FI +358 800 92066
FR +33 4 76 76 30 00
GR 800 11 71000
HU +36 80 630 106
IN +91 (22) 6697 4900
IT 800 089 300
NL 0800 022 05 85
NO +47 815 00 510
NZ 0800 80 22 99
PL +48 22 481 55 55
SE +46 (0)20-41 00 42
UK 0800 731 22 91
US 1-800-280-7801
ZA +27 11 504 4600
* The Bluetooth
®
word mark and logos are registered
trademarks owned by Bluetooth SIG, Inc. and any
use of such marks by Roche is under license.
Other trademarks and trade names are those of
their respective owners.
** Backup and loan pumps only.
Mainmenu
Press f to enter
Submenu
Press d to move forward …
SELECT USER MENU
ADVANCED / CUSTOM / STANDARD
BLUETOOTH ON / OFF
PAIRED DEVICE
ADD DEVICE
BOLUS
INCREMENT
PRIME AMOUNT
BASAL RATE LOCK
AUTOMATIC OFF
KEY LOCK
SOUND VOLUME
SIGNAL SETTINGS
ORIENTATION
LANGUAGE
DISPLAY CONTRAST
BOLUS DATA
ERROR DATA
DAILY TOTALS
TBR DATA
PUMP TIMER**
W1 CARTRIDGE LOW
Replace the cartridge as soon as possible.1.
W2 BATTERY LOW
Replace the battery within a few hours.1.
W3 REVIEW TIME
Check the date and the time and correct, 1.
if necessary.
W4 CALL FOR UPDATE
Contact your pump support for update. 1.
W5 PUMP TIMER**
Put the pump into RUN (if necessary).1.
Check MY DATA screen.2.
Contact your pump support.3.
W6 TBR CANCELLED
Put the pump into RUN (if necessary).1.
Ensure that this cancellation was 2.
intended.
Program a new temporary basal rate 3.
(if necessary).
W7 TBR OVER
Put the pump into RUN (if necessary).1.
W8 BOLUS CANCELLED
Ensure that this cancellation was intended.1.
Program a new bolus (if necessary).2.
W9 LOANTIME WARNING**
Put the pump into RUN (if necessary).1.
Check MY DATA screen.2.
Contact your pump support.3.
W10 BLUETOOTH FAULT
Redo pairing procedure for insulin pump 1.
and Meter.
E1 CARTRIDGE EMPTY
Change the cartridge immediately.1.
E2 BATTERY EMPTY
Change the battery immediately.1.
E3 AUTOMATIC OFF
Put the pump into RUN (if necessary).1.
E4 OCCLUSION
Check blood glucose level.1.
Disconnect the infusion set from body.2.
Exchange cartridge and infusion set.3.
Prime, connect the set to body and put 4.
the pump in RUN.
E5 END OF OPERATION**
1. Arrange immediate replacement of your
insulin pump.
2. Contact your healthcare professional
team if necessary.
E6 MECHANICAL ERROR
1. Disconnect the infusion set from body
and remove the cartridge.
2. Exchange the battery.
3. If possible activate the CARTRIDGE
CHANGE function.
4. Reinsert the cartridge and attach a new
infusion set.
5. Prime, connect the set to body and put
the pump in RUN.
6. If 3. not possible contact pump support.
E7 ELECTRONIC ERROR
1. Disconnect the infusion set from body.
2. Exchange the battery.
3. Prime, connect the set to body and put
the pump in RUN.
E8 POWER INTERRUPT
Check for correct time and date.1.
E9 END OF OPERATION**
Check MY DATA screen.1.
Contact your pump support.2.
E10 CARTRIDGE ERROR
Disconnect the infusion set from body and 1.
remove the cartridge.
Activate the CARTRIDGE CHANGE function.2.
Reinsert the cartridge and attach a new 3.
infusion set.
Prime, connect the set to body and put 4.
the pump in RUN.
E11 SET NOT PRIMED
Disconnect the infusion set from body.1.
Prime, connect the set to body and put the 2.
pump in RUN.
E12 DATA INTERRUPTED
Restart the data transfer.1.
REMINDER
Put the pump into RUN (if necessary).1.
E ND OF WARRANTY
Contact your pump support.1.
E13 LANGUAGE ERROR
Check Language.1.
Contact your pump support.2.