Alerts and errors
151
10.1.7 Alert A7: TBR OVER (Temporary Basal Rate over)
An alert A7: TBR OVER occurs when a temporary basal rate has ended.
Once the alert has been confirmed, the original basal rate (100%) will
resume automatically.
Press f twice to turn off and confirm the alert.
Decide whether a further temporary basal rate
change is appropriate and program, if necessary.
Note If desired, the alert A7: TBR OVER can be turned off using
Accu-Chek Insulin Pump Configuration Software. If turned
off, the alert does not occur when a temporary basal rate
ends.
10.1.8 Alert A8: BOLUS CANCELLED
If a bolus is cancelled during the start delay or after delivery has begun, an
alert A8: BOLUS CANCELLED occurs.
Press f twice to turn off and confirm the alert.
Put your insulin pump into RUN, if necessary.
Ensure that the cancellation was intended and,
if necessary, program a new bolus.
Note The actual bolus amount delivered prior to the cancellation
can be reviewed in the bolus history.
See the section 7.8.1 “Review the bolus history” for further
information.