Operation Manual

83
4
4.4 Programming a Multiwave Bolus
The Multiwave Bolus option is available in the ADVANCED user
menu. It is designed to better simulate the bodys insulin delivery,
and combines an immediate bolus delivery followed by an
Extended Bolus delivery. This option can be helpful when you
have meals that include both rapidly and slowly absorbed
carbohydrates.
The duration of bolus delivery can be programmed in 15-minute
intervals for up to 12 hours.
Note j The Multiwave Bolus is available only in the ADVANCED
user menu.
j You can add a Standard Bolus to an ongoing Multiwave
Bolus. If you cancel the Standard Bolus, the Multiwave
Bolus remains running.
j During the delivery of a Multiwave Bolus, the
programming of a second Multiwave Bolus or an
Extended Bolus is blocked. If you need a different
Multiwave Bolus, put your pump in STOP mode to
cancel the current bolus and program a new one.
When programming a Multiwave Bolus, make sure that you are
using the ADVANCED user menu and your pump is in RUN mode.
man_05504929001_H_en-us.indb 83 01.06.2016 08:14:48