Operation Manual
82
Note If a Temporary Basal Rate is active at the same time,
it is also cancelled when your pump is put into
STOP mode. The warning W8: BOLUS CANCELLED and
the warning W6: TBR CANCELLED are displayed.
Press f twice to confirm and turn off the first warning.
The second alert appears on the display. Press f twice
to confirm and turn off the second warning. Both alerts
are recorded in the warnings history.
See chapter 8 Warning W6: TBR CANCELLED and Warning
W8: BOLUS CANCELLED. Make sure that the cancellation
was deliberate and then program a new Extended Bolus
(and /or Temporary Basal Rate) if needed.
man_05504929001_H_en-us.indb 82 01.06.2016 08:14:48