Operation Manual

4
79
1. 2. 3.
Procedure for programming an Extended Bolus
Press d to move to the EXTENDED
BOLUS screen.
Press f to select. The BOLUS AMOUNT
screen appears.
Press a to increase or s to decrease
the bolus amount.
The duration of the last Extended Bolus
delivered appears.
Tip: You can program a Standard or Quick Bolus while an
Extended Bolus is ongoing. If you cancel the Standard Bolus, the
Extended Bolus continues to run.
During the delivery of an Extended Bolus, the programming of
another Extended Bolus or a Multiwave Bolus is blocked. If you
try to do this, you will receive the error message: “Bolus
currently running”.
If you want to program a different Extended Bolus, put the pump
in STOP mode to cancel the current bolus. Then put the pump in
RUN mode again to program a new Extended Bolus.
man_05504929001_H_en-us.indb 79 01.06.2016 08:14:48