Operation Manual
77
4
After this bolus delivery start delay, your pump beeps 3 times
and vibrates and begins to deliver the total bolus amount
programmed. The countdown of the remaining bolus amount
appears on the display.
4.2.4 Cancelling a Standard Bolus
During programming
(the bolus amount
blinks)
You have 3 options for cancelling during programming:
j If you do not press any key for 20 seconds after programming the bolus
amount, your pump returns to the RUN screen.
j Return to the Standard Bolus screen by pressing d + a simultaneously.
j Set the bolus amount to 0.0 units. Press f to exit.
No bolus is delivered.
During the start delay
(Standard Bolus symbol
blinks)
Press d + a, s or a. You hear a melody. Your pump returns to the RUN
screen. A warning W8: BOLUS CANCELLED occurs.
Press f twice to confirm and turn off the alert. See chapter 8 Warning W8:
BOLUS CANCELLED.
No bolus is delivered.
During bolus delivery
(countdown of bolus
amount)
Press and hold s or a for 3 seconds until you hear a melody.
You can view the actual bolus amount delivered before the cancellation by
looking at the bolus history. See section 5.4.1 Viewing your bolus history.
Bolus delivery is
interrupted.
man_05504929001_H_en-us.indb 77 01.06.2016 08:14:48