3
59
1. 2.
3.2.1 Short interruptions
3.
Press d to move to the STOP YOUR
PUMP screen.
Press f to put your pump in STOP
mode. Remove the infusion set or, when using a
disconnectable infusion set, disconnect
the transfer set from the headset and use
the protective covers that came with it.
Resuming insulin delivery
Reconnect your infusion set or insert a
new one and put your pump into RUN
mode.
1.
Where appropriate, attach a new
Accu-Chek infusion set. (See section
3.4 Changing the infusion set).
Measure your blood glucose level within
2 hours of restarting pump therapy to
check the function of the system and to
ensure insulin delivery.