Operation Manual

54
Note Your pump cannot detect infusion set leakage. If you
detect any loss of insulin and all parts are correctly
tightened, immediately replace the leaking component.
Because insulin delivery has been interrupted, check your
blood glucose level and take appropriate action according
to your doctor’s or healthcare team’s instructions.
Note If the battery is too low, the start-up process may not be
able to complete. In this case, the pump beeps five tones
and keeps restarting the process until you remove the
battery or until it is empty.
Note Check the amount of insulin remaining in the cartridge at
least once a day. If you press f from the RUN or STOP
screen, your pump displays the remaining cartridge
content in units.
Before going to sleep, make sure that:
j the cartridge contains enough insulin to last through
the night
j the time and date are set correctly