Operation Manual
51
2
2.11 Stopping insulin delivery
For your safety, you must only perform certain functions and
actions when the pump is in STOP mode. These functions and
actions include:
j changing the cartridge
j connecting and disconnecting the adapter and / or infusion set
j priming the infusion set
j data transfer
c CAUTION
When your Accu-Chek Spirit Combo insulin pump is in STOP
mode, no insulin is delivered to your body. You must put your
pump into RUN mode to continue insulin delivery.
Procedure to stop insulin delivery
Press d to move to the STOP YOUR
PUMP screen.
1. 2.
Press f to select. Your pump returns to
the STOP screen.
Your pump is now in STOP mode.
No insulin is being delivered.
man_05504929001_H_en-us.indb 51 01.06.2016 08:14:41