Operation Manual

42
19.
Press f after correctly inserting the
cartridge.
Your pump performs a self-test. See
section 2.3 Start-up process.
When the self-test process is finished,
the INFUSION SET PRIMING screen is
displayed.
See section 2.8 Priming the infusion set. Start from step 4 in the
Procedure for priming the infusion set.
c CAUTION
If an empty cartridge is inserted into the pump (e.g. for training
purposes), the pump displays the programmed insulin deliveries
(infusion set priming, basal rate and bolus deliveries), even
though no insulin is being delivered due to the empty cartridge.
20.
Check that the infusion set luer-lock
connector is still correctly connected to
the adapter. Tighten the infusion set
clockwise by hand, until it sits tightly in
the adapter.
If the end plate of the piston rod is not
correctly aligned with the cartridge
plunger, remove the cartridge from the
cartridge compartment and repeat steps
9 to 15.
17.
18.
man_05504929001_H_en-us.indb 42 01.06.2016 08:14:38