Operation Manual

2
41
The cartridge is correctly inserted when
the end plate of the piston rod is correctly
seated against the cartridge plunger.
16.
Position your pump diagonally downwards
(with the adapter pointing down). Insert
the cartridge into the cartridge
compartment.
It is not necessary to apply pressure.
Do not over-tighten the adapter,
otherwise it will be difficult to remove the
cartridge, adapter and cartridge plunger,
and you could cause damage to your
pump.
Do not insert the cartridge into your
insulin pump without first connecting the
infusion set tubing. If the tubing is not
connected first, insulin may flow into the
cartridge compartment, which may result
in insulin entering your pump and causing
damage to your pump.
Twist (do not push) the adapter clockwise,
until it is correctly seated against the
cartridge compartment.
Do not hold your pump upright when
inserting the cartridge. Holding the pump
upright may lead to insulin flowing into
the cartridge compartment, which may
block the piston rod and potentially cause
an error.
15.
13. 14.