Operation Manual

2
39
8.
Make sure your infusion set is
disconnected and the cartridge has been
removed.
Press and hold f for 3 seconds, until
you hear a melody.
The piston rod begins to return. Do not
touch or otherwise disturb the piston rod
while it is rewinding, as this could result
in damage to your pump.
7.
6.
When the piston rod has returned
completely, the CARTRIDGE VOLUME
screen displays the maximum volume,
315 U.
Press d to move to the CARTRIDGE
CHANGE screen.
Press f to select.
5.
Hold the pump with the adapter facing the
ground. Remove the cartridge, if
necessary. Unscrew carefully. Do not pull
the cartridge out until it is completely
unscrewed.
c CAUTION
Do not change the cartridge if the
cartridge compartment is not
completely dry. When the piston rod is
rewinding, insulin could enter your
pump and lead to malfunction. If
necessary, use a soft cloth to dry the
compartment.
4.