c CAUTION
To prevent water from entering the pump casing, only insert or
change the battery in a dry environment. Make sure that the
battery cover seal is not worn or missing, and that the battery
is correctly inserted.
Note If your pump was in RUN mode before you removed the
battery, the error E8: POWER INTERRUPT will occur when
the new battery is inserted. Press f twice to confirm
and snooze the error. Check the time and date and correct
them if necessary. Then put your pump into RUN mode.
See section 2.4 Setting the time and date.
w WARNING
It is important that you always check that the date and time
are set up correctly before you put your pump into RUN mode
after a battery change. Otherwise there is a risk that too
much or not enough insulin will be delivered because of a
shift of basal rate time blocks. Basal rates that are delivered
at the wrong time can lead to hyper- or hypoglycemia.