Appendix B: Configuration parameters
Typical range available on
the pump and standard
parameter settings*
Maximum parameter
range adjustable with the
appropriate Accu-Chek PC
software
Basal rate
Hourly basal rate + max. TBR
(hourly basal rate combined with temporary
basal rate, max. increase)
62.5 U / h max . amount : 0–25 0.0 U / h
Default basal rate 0 U/h 050 U/h
Hourly basal rate increment 0.01 U/h (up to 1.00 U / h),
0.05 U/h (up to 10.0 U / h) and
0.1 U/h (up to 25.0 U / h) 0.0 U/h, 0.05 –25.0 U / h
Hourly basal rates 25.0 U / h max . amount : 0–25.0 U / h
Temporary Basal Rate
TBR increases or decrease in increments of
TBR duration
TBR duration increments
0–250 %
10 %
max. 24 h
15 min
0500 %
10 %
15 min–50 h
15 min, 30 min, 1 h
Basal rate profiles available
Basal rate lock
5
off
1–5
on or off
* Note: These factory settings may vary in some countries and depend on the
changes you, your doctor or healthcare team make using your insulin pump or
pump programming software for healthcare professionals.