Operation Manual
184
Start-up graph over the stabilization period
The start-up graph shows changes in the flow rate over the
stabilization time.
flow rate (ml / h)
1455
1485
1515
1545
1575
1605
1635
1665
1695
1725
1755
1785
1815
1845
1875
1905
1935
1965
1995
2025
2055
2085
2115
2145
2175
2205
2235
2265
2295
2325
2355
2385
2415
2445
2475
2505
2535
2565
2595
2625
2655
2685
2715
2745
2775
2805
2835
2865
2895
2925
0.002
0.000
–0.002
time (min)
real flow
target flow
0.01 ml is equivalent
to 1 U (U100 insulin)
Bolus
For U100 insulin the maximum deviation of
a maximum bolus (25 U and 50 U) is ≤ ±5 % and
the maximum deviation of a minimum bolus (0.1 U) is ≤ ±30 %*.
* The measurements were made according to IEC 60601-2-24:2012 with an Accu-Chek Rapid-D Link infusion set 12 / 100 at room temperature.