Operation Manual
183
A
Accuracy of flow rate for the Accu-Chek 3.15 ml plastic
cartridges and the Accu-Chek Spirit 3.15 ml cartridge
system at minimum hourly basal rate of 0,05 U/h
Trumpet curve plotted from data after the end of the stabilization
period.
The trumpet curve shows the accuracy of the delivery rate in
relation to the observation period.
percentage error of flow rate (%)
50
100
–
150
–250
–300
–350
060 120 180 300 660 780 960240 360 540 600 840
–50
–
100
420 480 720 900
–
200
0
– 292.0000
–86.0000
–39.2000
– 20.7273
– 12.4211 – 8.0000
68.0000
22.0000
11.2000
5.0909
2.3158
observation window duration (min)
minimum measured variation
maximum measured variation
overall percentage error
The maximum deviation of the delivered amount (overall mean
percentage error of flow rate) for 100U insulin is ≤ 30%*.
It is in the responsibility of the health care professional to
decide whether the accuracy of the flow rate is sufficient
for the according patient.
* The measurements were made at the minimum rate of 0,05U/h according to IEC 60601-2-24:2012 with an Accu-Chek Tender infusion set PC 17 / 110 at room
temperature.