Technical standards
Table 201: Electromagnetic emission
Guidance and manufacturer’s declaration – electromagnetic emissions
The Accu-Chek Spirit Combo insulin pump is intended for use in the electromagnetic environment specified below. The customer or
the user of the Accu-Chek Spirit Combo insulin pump should ensure that it is used in such an environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The Accu-Chek Spirit Combo insulin pump must emit electro-
magnetic energy in order to perform its intended function. In rare
cases, nearby electronic equipment may be affected.
RF emissions
CISPR 11
Class B The Accu-Chek Spirit Combo insulin pump is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable –
Voltage fluctuatio ns / f licker
emissions
IEC 61000-3-3
Not applicable –