Operation Manual

173
A
General Technical Data
Infusion sets The Accu-Chek infusion sets with a luer-lock connector (see Appendix C).
Data transfer between pump and a PC Infrared interface
Safety system The safety concept is based on a control system which consists of two microprocessors
(1main processor and 1 supervisor processor). Whenever a defect or fault occurs in the
main processor, it is identified by the supervisor processor. If, for example, overdelivery
occurs, the motor is immediately switched off and the error E7: ELECTRONIC ERROR
occurs. The main processor and supervising processor signal the error E7 with beeps and
vibrations. In addtion, the main processor shows the specific error E7 number on the pump
display. Conversely, the main processor can also determine at any time whether the
supervisor processor is working correctly.
IPX8 Protected against the effects of temporary immersion in water under standardized
conditions (up to 60 minutes and 2.5 meters [8 feet]).
Electric shock protection The Accu-Chek Spirit Combo insulin pump and the Accu-Chek infusion sets are applied
parts of the type BF.
* See the instructions for use of the insulin you are using for information on the acceptable temperature range for operation, storage and handling.
** Determined using measurement methods according to IEC 60601-2-24-1998.
man_05504929001_H_en-us.indb 173 01.06.2016 08:14:52