Operation Manual
172
General Technical Data
Maximum time before the error
E4: OCCLUSION** occurs
Plastic cartridges:
j at medium basal rate 1.0 U / h: ≤ 2 h
j at minimum basal rate 0.05 U / h: ≤ 24 h
Maximum missing volume before the
error E4: OCCLUSION** occurs
Plastic cartridges: ≤ 2.0 U
Maximum system occlusion alarm
threshold
< 400 kPa (4.0 bar)
Maximum alarm threshold < 400 kPa (4.0 bar)
Maximum pressure 400 kPa (4.0 bar), for plastic cartridges and infusion sets.
Flow (delivery rate) During infusion set priming and bolus: 0.2 U / sec.
Maximum quantity delivered at
a single fault condition
≤ 1.0 U
Cartridge The Accu-Chek Spirit 3.15 ml cartridge system and the Accu-Chek 3.15 ml plastic cartridge
with a luer-lock connection. Sanofi Aventis Insuman Infusat cartridges*** have been
evaluated and approved for use with the Accu-Chek Spirit Combo insulin pump.