Problem Recommended Action
There are air bubbles in the cartridge and /or infusion set. Check your blood glucose level.
In the cartridge: Disconnect the infusion set from the headset and
re-prime the infusion set. If the air bubbles remain, replace the
cartridge. Always fill the cartridge with insulin that is at room
temperature.
In the infusion set: Disconnect from the headset and prime your infusion
set. Reconnect the headset and ensure that all connections are tight and
secure.
The cartridge is empty. Replace the cartridge. (Cartridges are for single use only.)
The infusion set is disconnected or dislodged. Check your blood glucose level. Make sure that all the connections are
tight. Inspect the connections for leaking insulin. Change your infusion
set and infusion site.