Error E9: END OF OPERATION (loan pumps only)
Your pump timer has reached zero. The pump changes to STOP
mode and will no longer operate. The error E9: END OF
OPERATION appears on the display. Before
this error occurs, the warning W9: LOANTIME WARNING will
remind you to replace your pump soon. If necessary, use the
alternative therapy plan provided by your doctor or healthcare
team.
j Press f twice to snooze and to confirm the error.
j Contact your Accu-Chek Customer Care (details on inside back
cover) to discuss the options for further use of the Accu-Chek
Spirit Combo insulin pump.
Note This setting is country-specific and may not be visible on
your pump.
The warning W9: LOANTIME WARNING and the error E9:
END OF OPERATION are based on running time (the total
time the pump is in RUN mode) and not on calendar
years.
Error E10: CARTRIDGE ERROR
The CARTRIDGE CHANGE function was not correctly performed.
j Press f twice to snooze and to confirm the error.
j Remove or disconnect the infusion set from your infusion site.
j Remove the cartridge.
j Reinsert the cartridge. (See section 2.7 Inserting the cartridge)