Operation Manual

151
8
Warning W7: TBR OVER (Temporary Basal Rate Over)
A Temporary Basal Rate has ended. The original basal rate
(100 %) will resume automatically once the warning has
occurred.
j Press f twice to snooze and confirm the warning.
j Decide whether a further Temporary Basal Rate change is
required. If necessary, program a new Temporary Basal Rate.
Warning W8: BOLUS CANCELLED
A bolus was cancelled during the start delay or after delivery
began.
j Press f twice to snooze and to confirm the warning.
j If necessary, put your pump in RUN mode.
j Make sure that the cancellation was intended and then
program a new bolus if needed.
Note The actual bolus amount that was delivered before the
cancellation can be reviewed in the bolus history. See
section 5.4.1 Viewing your bolus history.
man_05504929001_H_en-us.indb 151 01.06.2016 08:14:51