90
Note During Temporary Basal Rate delivery, the increase or
decrease in percentage, the increased or decreased
hourly basal rate (hourly basal rate plus TBR percentage),
and the remaining time are displayed in the RUN screen.
An arrow pointing upward symbolizes a Temporary Basal
Rate increase; an arrow pointing downward symbolizes a
Temporary Basal Rate decrease.
Note If an Extended Bolus or a Multiwave Bolus is active at the
same time, the remaining time and amount of the bolus
and the amount of the increased or decreased hourly
basal rate are displayed in the RUN screen.
Example: Extended Bolus running
Remaining duration
Extended Bolus symbol
TBR OVER Alert
At the end of a Temporary Basal Rate, the warning W7: TBR
OVER is displayed. Press f twice to confirm and turn off the
warning. See chapter 8 Warning W7: TBR OVER.