Operation Manual

5
29
Meter Memory, Setup, and Data Transfer
Turn the meter on.
The flashing test strip
symbol appears on the
display.
Press and hold
until
set-up appears on the
display.
1
Press and release
repeatedly until set-up,
and the flashing 1Hr or
2Hr appear.
2
Press and release or
to select 1Hr or 2Hr.
Press and release
to
set the option.
To set up more options,
press and release
.
To exit, press and hold
until the flashing test
strip symbol appears on
the display.
3
This display appears when
a post-meal test reminder
occurs.
4
NOTE
The post-meal test reminder:
beeps 1 or 2 hours after the blood glucose test to
remind you to do a post-meal test.
beeps every 2 minutes up to 3 times.
turns o by inserting a test strip or pressing any
button.
If a test was performed within 15 minutes of a test
reminder, the test reminder does not beep.
If the meter is on at the test reminder time, the test
reminder does not beep.
Exposure to cold conditions may disable test
reminders until the meter is turned on.
Setting the Post-Meal Test Reminder