Performing a blood glucose test
4
Cause Action
The blood drop touched the test area for too long.
Remove your finger from the test area as soon as
the beep tone sounds and Test in progress is
displayed.
The blood drop was used for a second test. Use a new, fresh blood drop for each test.
The temperature was too low or too high during
the test.
For blood glucose tests, the permitted
temperature range is between +10 and +40°C.
Move to a place where the temperature is at least
+10°C and at most +40°C and wait for the
temperature of the meter to adjust to this
temperature.
The meter or test cassette was stored incorrectly.
Store the meter and test cassettes according to
the specified storage conditions (see Testing and
storage conditions page 117 and the package
insert for the test cassette).
A source of error described in the package insert
for the test cassette applies.
Check whether one of the mentioned sources of
error applies.
The meter was dropped. Perform a control test.
W
If your meter was dropped, this may also lead to implausible test results or error messages. Contact the
customer support and service centre.