130
Technical data
16
Technical data
Meter type
Accu-Chek Mobile
Model U1
Catalogue no./serial
no.
See type plate on the back of the meter
Test principle
Determination of glucose in fresh capillary blood by reflectance photometry.
When using dierent samples, refer to the package insert of the Accu-Chek
Mobile test cassette.
Your meter displays blood glucose values that refer to plasma although you
always apply whole blood to the test area.
You can find information on how the test works and on the referenced test
principle in the package insert of the Accu-Chek Mobile test cassette.
Measuring interval See the package insert of the Accu-Chek Mobile test cassette
Blood volume See the package insert of the Accu-Chek Mobile test cassette
Measuring time Approx. 5seconds (depending on the concentration)
Power supply
2 alkaline-manganese batteries (1.5 V; type AAA, LR 03, AM 4 or micro) or
2 rechargeable NiMH batteries (type AAA), backup battery: 3-volt lithium coin
cell type CR1025
Battery life
With the enclosed batteries: approximately 500tests or approximately 1year
(less if the brightness of the screen is set to level3 or the volume is set to
level4 or 5, or in acoustic mode due to the higher power consumption)
Automatic power o After 1 or 2minute(s), depending on the operating status