Operation Manual

127
14
Messages and problems
E-6E-5
Move into the shade or shield the
meter, for example, with your
body.
Repeat the test.
You...
• have not washed your fingers or they are soiled or sticky.
• applied a contaminated blood drop (for example, due to food, drink
or glucose residue on your finger).
• pressed your finger onto the test area.
• moved the tape of the test cassette.
• smeared the blood on the test area.
• did not keep your finger still or
• did not remove your finger from the test cassette immediately
after the beep tone.
Wash your hands with warm water and soap and rinse them well.
Dry your hands thoroughly with a clean towel before obtaining blood.
Touch the blood drop to the test area, which is located between the
guidance tabs at the tip of the cassette. Place your finger lightly on
the guidance tabs without pressing it onto the test area between
them. Only the blood drop should touch the test area of the cassette.
Keep your finger as still as possible.
Apply the blood drop or control solution only to the centre of the test
area without pressing on the test area.
Remove your finger from the test cassette or remove the brush with
control solution from the test area as soon as the beep tone sounds
and Test in progress is displayed. The beep tone helps you to obtain
a reliable test result.
See the next page for a further
E-6 message.
man_08252963001_02_EN-AUS.indb 127 27.02.2018 10:12:51