Operation Manual
123
14
Messages and problems
Problems
Problem and possible causes Solution to the problem
The meter will not turn on.
The batteries are almost empty, empty or none
are inserted.
Insert new batteries.
The batteries were inserted the wrong way.
Remove the batteries and re-insert them as shown
in the battery compartment.
The ambient temperature is low and the
batteries are almost empty.
Move to a place where the temperature is between
+10 and +40°C and wait for the temperature of
the meter to adjust to this temperature. Insert new
batteries.
Condensation has made the electronics damp. Allow the meter time to dry slowly.
The meter is defective.
Contact the customer support and service centre
(see page 134).
The meter displays dashes instead of the time and date.
The meter has no longer saved the time and
date.
The meter was stored without batteries and the
built-in backup battery is empty.
Reset the time and date.
The meter is turned on, but the display remains blank.
The display is defective.
Contact the customer support and service centre
(see page 134).
man_08252963001_02_EN-AUS.indb 123 27.02.2018 10:12:51