Operation Manual

122
Messages and problems
14
This message appears at the
start of a test if the ambient
temperature or the temperature
of the meter is between +8 and
+10°C 1 or between +40 and
+42°C 2.
Use
to select whether you
want to Test or Quit.
Press
.
• Test: The meter advances a
test area.
• Quit: The meter opens the
Main menu. Close the tip
cover.
If you decide to perform the test
in spite of the warning, the test
result is flagged with the
symbol
and stored.
W
Do not use blood glucose results obtained despite this warning as a
basis for making therapeutic decisions. These test results may be
incorrect. Incorrect test results can cause the wrong therapy
recommendation to be made and thus produce serious adverse
health eects. Move to a place where the temperature is between
+10 and +40°C and wait for the temperature of the meter to
adjust to this temperature.
Temperature
12