Operation Manual
113
11
Cleaning the blood glucose monitoring system
3
3
Dry the finger pricker and cap
with a dry cloth.
Remove the lancet drum from
the finger pricker.
2
2
Wipe the outside and the inside
of the cap thoroughly (see
Disinfecting).
Remove the cap from the finger
pricker.
1
1
Slightly dampen a cloth with
water.
Wipe the outside of the finger
pricker thoroughly (see
Disinfecting).
Remove the finger pricker from
the meter.
Disinfecting
3
Wrap the whole finger pricker in
the cloth.
Leave the finger pricker wrapped
up for 8minutes.
Unwrap the finger pricker and let
it dry in the air.
2
Wipe the outside and the inside
of the cap thoroughly for
2minutes.
Place the cap back on again.
1
Dampen a cloth well with 70%
isopropanol.
Wipe the outside of the finger
pricker thoroughly for 2 minutes.
Cleaning
Preparation
man_08252963001_02_EN-AUS.indb 113 27.02.2018 10:12:47