Cleaning the blood glucose monitoring system
9
Re-insert the test cassette in the
meter.
kim_manual_0800
8
Open the tip cover again.
kim_manual_0795
kim_manual_0795
7
Carefully dab the interior of the
tip cover with a slightly
dampened cotton swab.
Remove any residues left by the
cotton swab.
Allow the meter to dry
thoroughly.
Close the tip cover.Close the cassette compartment
cover.
Press the cassette compartment
cover closed.
The cassette compartment cover
must close with an audible
CLICK.
H
Make sure that no liquid enters the meter. Avoid scratching the measuring optics.
CLICK